Technical data - SpectroShade™ "Office" Derma
route: dermatological division -> office data
Type of instrument:
- Spectrophotometer
Spectral data:
- By mean of a diffraction grating
- Spectral Interval: from 450nm to 980nm
- Spectral Resolution: 30nm
Measurement data:
- Calibrated Data: from 483nm to 950nm
- Steps: 33nm
- Illumination 2 x 45°, monochromatic, telecentric
- Reading at 0° / telecentric
- Reading area: ~18x14mm on 640x480 points
- Number of points: 640x480 = 307.200 (= N° of spectral curves in every acquisition)
- Resolution: ~0,03x0,03mm for each point
Sensors:
- 2 CCDs, progressive scanning, 640x480 points
- CCD 1: colors sensors (with mosaic gradations for color images)
- CCD 2: black and white sensor (for spectrum data reading)
Selling Kit:
- PC Windows 2000 or more (not included)
- CPU: Pentium III 800 MHz
- 128 Mbyte RAM
- 20 Gbyte HD
- 1.44 MByte Floppy Drive
- CD / CD-R / CD-RW writer
- LAN / Modem
Standards:
- Power voltage: 115V 60Hz / 230V 50Hz (automatic)
- Power consumption: 450 VA (fuses 8A / 4A delayed)
- Environment temperature: 0°C - 30°C
- Air humidity: 30% - 80%
- Dimensions: (WxHxD) 19 x 45 x 42 cm
- Weight: 18 Kg
Minimum computer requirements:
- Safety: EN 60601-1:1990 + A1:1993 + A2:1993 + A13:1996
- EMC: EN 60601-1-2:1993 + EN 61326-1:1997 + EN 55022:1994 + CISPR 22:1997
- EU Dir.: 93/68/CEE + 89/336/CEE + 73/23/CEE 93/42/CEE Medical Devices - Class I